3 Assembly

872 Extension Module IC Pump

■■■■■■■■

17

3 Mounting the bottle holder

1

2

1

2

Figure 9

Mounting the bottle holder

1

Cover stoppers

2

Cylinder screws

■

Place the bottle holder on the extension module.

■

Tighten the cylinder screws (9-2) with an 6.2621.100 3 mm hexa-
gon key.

■

Insert covering stoppers (9-1).

4 Connecting the extension module

■

Plug a 6.2156.060 cable into the connector In of the extension
module and screw it tight.

■

Plug the other end of the cable into the Extension module con-
nector of the IC instrument and screw it tight.

5 Optional: Connecting a further extension module

■

Plug the cable 6.2156.060 or a longer 6.2156.070 cable (optional
accessory) into the connector In of the second extension module
and screw it tight.

■

Plug the other end of the cable into the Out connector of the first
extension module and screw it tight.