3 Operation

709 IC Pump

22

3.4.2

Maximum pressure limit

The maximum shutoff pressure of the pump can be set in steps of

0.1 MPa from 0.1

50 MPa with the pump drive switched on or off. Pro-

ceed as follows:

1 Select "P

max

"

Press <SELECT> key 4 until the "P

max

" LED 12 lights up.

2 Set shutoff pressure

Press <

> key 5 or <

> key 6 until the desired shutoff

pressure (in MPa) appears in display 1.

3 Store shutoff pressure

Release <

> key 5 or <

> key 6: the value shown in display

1 at the time of release will be stored.

The set limit value should lie between 5.0 and 10.0 MPa above the par-

ticular operating pressure or the maximum admissible operating pres-

sure of the column. If the pump exceeds the preset limit value during

operation, the pump drive is switched off within 1 pump cycle and dis-

abled. If the pump has been switched off by the maximum pressure

limit monitoring, the LED 12 flashes, the shutoff pressure can be re-

quested in the display "P

actual

" (see section. 3.7).

Pressing the <R/S> key 3 re-enables the pump drive, LED 12 no

longer flashes and the pump is in the standby mode. Pressing the

<R/S> key 3 again puts the pump back into operation.

The 2.709.0110 IC Pump MF (metal-free version) should only be

operated in the pressure range of 0

25.0 MPa (0

250 bar). With

higher pressure, the 6.2824.100 PEEK pump head can be damaged.

So make sure that the maximum shutoff pressure

P

max

of the pump is

always set to

25.0 MPa.

3.5

Switching the pump drive on and off

Switching on Pressing the <R/S> key 3 switches on the pump drive.

While the pump drive is running, LED 2 above the

<R/S> key 3 lights up.

Switching off To switch off the pump drive, press the <R/S> key 3

again. LED 2 goes out.