3 Operation

709 IC Pump

22

3.4.2

Maximum pressure limit

The maximum shutoff pressure of the pump can be set in steps of

0.1 MPa from 0.1

…

50 MPa with the pump drive switched on or off. Pro-

ceed as follows:

1 Select "P

max

"

Press <SELECT> key 4 until the "P

max

" LED 12 lights up.

2 Set shutoff pressure

Press <

↓

> key 5 or <

↑

> key 6 until the desired shutoff

pressure (in MPa) appears in display 1.

3 Store shutoff pressure

Release <

↓

> key 5 or <

↑

> key 6: the value shown in display

1 at the time of release will be stored.

The set limit value should lie between 5.0 and 10.0 MPa above the par-

ticular operating pressure or the maximum admissible operating pres-

sure of the column. If the pump exceeds the preset limit value during

operation, the pump drive is switched off within 1 pump cycle and dis-

abled. If the pump has been switched off by the maximum pressure

limit monitoring, the LED 12 flashes, the shutoff pressure can be re-

quested in the display "P

actual

" (see section. 3.7).

Pressing the <R/S> key 3 re-enables the pump drive, LED 12 no

longer flashes and the pump is in the standby mode. Pressing the

<R/S> key 3 again puts the pump back into operation.

The 2.709.0110 IC Pump MF (metal-free version) should only be

operated in the pressure range of 0

…

25.0 MPa (0

…

250 bar). With

higher pressure, the 6.2824.100 PEEK pump head can be damaged.

So make sure that the maximum shutoff pressure

P

max

of the pump is

always set to

≤

25.0 MPa.

3.5

Switching the pump drive on and off

Switching on Pressing the <R/S> key 3 switches on the pump drive.

While the pump drive is running, LED 2 above the

<R/S> key 3 lights up.

Switching off To switch off the pump drive, press the <R/S> key 3

again. LED 2 goes out.