7 remote control commands – Metrohm 709 IC Pump User Manual

Page 50

Advertising
background image

5 Interfaces

709 IC Pump

42

5.1.7

Remote control commands

Object Meaning

Entry

range

&

Root

│

├

.

Pump

Pump

│

├

.

Start

Start the pump

$G, $S

│

├

.

Soft

Soft start the pump (within 5 s)

$G, $S

│

└

.

Identifier

Instrument identification

up to 8 ASCII characters

│

├

.

Parameter

Parameters

│

├

.

A

Metrohm pump head

│

│

├

.

Flow

Flow rate

0.01

…

5.00 mL/min

│

│

├

.

Pmax

Upper shutoff pressure

P

min

…

50.0 MPa

│

│

├

.

Pmin

Lower shutoff pressure

0.1

…

P

max

MPa

│

│

└

.

NX

Correction factor

0.90...1.10

│

├

.

P

Parameter for prep. head

│

│

├

.

Flow

Flow rate

read only

│

│

├

.

Pmax

Upper shutoff pressure

read only

│

│

├

.

Pmin

Lower shutoff pressure

read only

│

│

└

.

NX

Correction factor

read only

│

├

.

M

Parameter for micro head

│

│

├

.

Flow

Flow rate

read only

│

│

├

.

Prnax

Upper shutoff pressure

read only

│

│

├

.

Pmin

Lower shutoff pressure

read only

│

│

└

.

NX

Correction factor

read only

│

└

.

R

Parameter for reserve head

│

├

.

Flow

Flow rate

0.01...5.00 mL/min

│

├

.

Pmax

Upper shutoff pressure

0.1

…

50.0 MPa

│

├

.

Pmin

Lower shutoff pressure

0.1

…

50.0 MPa

│

└

.

NX

Correction factor

0.90

…

1.10

│

├

.

Info

Current configuration

│

├

.

Head

Pump head

│

└

.

Actual

Current data

│

├

.

Pressure

Pressure

│

├

.

Flow

Flow rate

│

├

.

Pmax

Upper shutoff pressure

│

├

.

Pmin

Lower shutoff pressure

│

├

.

NX

Correction factor

│

├

.

MaxPmax

Maximum possible upper shutoff pressure

│

├

.

MinPmax

Minimum possible upper shutoff pressure

│

├

.

MaxPmin

Maximum possible lower shutoff pressure

│

└

.

MinPmin

Minimum possible lower shutoff pressure

│

└

.

Setup

Instrument settings

├

.

Trace

Message if values or paths change

ON,OFF

└

.

Autoinfo

Automatic message on changes:

├

.

P

when power switched on

ON,OFF

├

.

R

when instrument in "ready" status

ON,OFF

├

.

S

when instrument in "stop" status

ON,OFF

├

.

E

for error message

ON,OFF

└

.

Pressure

on every change of measured pressure value

ON,OFF