Figure 18, Working range - disposal, Figure 19 – Metrohm Liquid Handling Station User Manual

Page 25: Working range - dilution unit

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

■■■■■■■■

17

3 If necessary, correct the angle value.

4 Move to External position 2 (16-2).

5 Check whether the sample tube is centered over the "Rinsing unit -

Disposal" position.

Figure 18

Working range - Disposal

6 If necessary, correct the angle value.

7 Move to External position 3 + 4 (16-3), (16-4).

8 Check whether the sample tube is centered precisely over the open-

ing of the dilution unit.

Figure 19

Working range - Dilution unit

9 If necessary, rotate the cover of the mixing vessel until the opening is

centered under the sample tube.

10 If necessary, correct the angle value.