Spikes / signal jump in voltammogram, Oxygen interference – Metrohm 797 VA Computrace User Manual

Page 280

Advertising
background image

9 Troubleshooting

797 VA Computrace – Software

268

2. If organic components are present: carry out a UV digestion or

other suitable sample preparation.

3. Concentration of the analyte is too high: dilute.
4. Electrolyte solution too old: make up a new one. Its working

life with organic additives may be as short as 1 day or less.

With HMDE:
1. Standard addition solution containing metal complexing agents

need time to form the metal complex.

Spikes / signal jump in voltammogram

1. For MME: check the electrode.
2. Reduce the dynamic range of the potentiostat (see Potentio-

stat, section 3.3).

Oxygen interference

Oxygen can be electrochemically reduced and produces two waves
in the voltammogram, one of which is characterized by the ap-
pearance of a pronounced maximum. The oxygen reduction can
interfere for two reasons:

•

The signals of the analytes are masked by the oxygen waves.
This becomes noticeable primarily in trace analysis as the oxy-
gen is present in a relatively high concentration in solutions
saturated with air (ca. 8 mg/L at room temperature).

•

The hydrogen peroxide formed in the first step of the oxygen
reduction can react further with certain substances.

For these reasons, oxygen must be removed from the analysis solu-
tion before the polarographic analysis by saturation with inert gas
(usually nitrogen). With the inert gas flow rate of ca. 20 L/h set on
the 797 VA Computrace stand in the factory, a purging time of
3...5 min usually suffices.

Compare the curves (0.1 mol/L KNO3) before and after purging:

Before purging (Still oxygen in the solution):