3.2 VA measurement modes

797 VA Computrace – Software

43

Sweep parameters:

Hydrodynamic (measurement) [ on, off ; off ]

Enable/disable stirring of the RDE/SSE during the sweep.

Start potential (V)

[ -5...+5 V ; -0.9 V ]

Start voltage for the voltage sweep.

End potential (V)

[ -5...+5 V ; -0.1 V ]

Final voltage for the voltage sweep.

Voltage step (V)

[ > 0 V ; 0.006 V ]

Voltage step for direct voltage ramp.

Amplitude (V)

[ > 0...+1 V ; 0.02 V ]

Voltage amplitude of the square wave voltage superimposed
on the direct voltage.

Frequency (Hz)

[ > 0...2000 Hz ; 50 Hz ]

Frequency of the superimposed square wave voltage, which
defines the voltage step time (

Voltage step time = 1 / Fre-

quency

).

Sweep rate (V/s)

[ read only ]

Display of the ramp slope calculated as

Voltage step × Fre-

quency

.

Notes:
• The SqW measurement mode can only be used with HMDE or

RDE electrodes.

•

If your curves have high noise, put

Highest current range

and

Lowest current range

in the

POTENTIOSTAT

window on the

same level (see section 9.6

SqW Problems

)

• The following condition applies to the definition of the

Voltage

step time

:

Frequency ≤ 2000 Hz ⇒ Voltage step time ≥ 500 µs

• The measurement time

t (i)

is defined as follows:

Voltage step time ≥ 80 ms

→

t (i) = 20/16.67 ms

(power fre-

quency 50/60 Hz)

Voltage step time < 80 ms

→

t (i) = 0.5 × Voltage step time