VICI 340 Dynacalibrator User Manual

Page 6

Advertising
background image

2

Introduction

As indicated in Figure 1, the carrier stream passes through scrubber #1 and
30-micron line filter LF1 to the differential pres sure regulator DPR1. The scrubber
contains spe cially-activated charcoal designed for broad-based scrubbing action.
This scrubbing medium is particularly effective in removing sulfur dioxide and
other sulfur compounds.

The differential pressure regulator and carrier flow restrictor FR1 together
provide a highly stable carrier stream flow rate to the permeation chamber.
The standard carrier flow restrictor consists of a disk with a 0.006" diameter
orifice. The flow rate through the orifice is a function of the pressure differential
across it and this differential is maintained at 3.2 psi by DPR1.

The output of the permeation chamber is fed to a mixing tee, where the calibrat-
ed dilution stream is added to set the final concentration. The combined stream
is then fed through another tee to the rear panel overflow vent and span outlet.

SPAN/STREAM

OUTLET

DILUTION

STREAM

INPUT

TEE

TEE

MIXING

TEE

OVERFLOW

VENT

CARRIER

STREAM

INPUT

SCRUBBER

#1

SCRUBBER

#2

LINE FILTER

LF1

DIFFERENTIAL

PRESSURE

REGULATOR

DPR1

CARRIER

FLOW

RESTRICTOR

FR1

PERMEATION

CHAMBER

CHAMBER

FLOWMETER

FM1

DILUTION

FLOWMETER

FM2

TEE

DIFFERENTIAL

PRESSURE

REGULATOR

DPR2

DILUTION FLOW

CONTROL VALVE

FCV2

SOLENOID

CONTROL

VALVE

SV1

CHAMBER

VENT

NO

NC

C

Figure 1: Model 340 plumbing schematic

For those analyzers sensitive to sample feed pressure, the overflow vent dumps
excess calibration gas to ensure that the analyzers receive span gas at near at-
mospheric pressure. The overflow vent is usually left open, but can be plumbed
to an external exhaust point. Any exhaust plumbing must be sized so as not to
increase pressure into the analyzer under calibration.

Referring again to Figure 1, note that the dilution stream passes through
scrubber #2 and feeds into differential pressure regulator DPR2. The differential
pressure regulator maintains a fixed pressure differential across dilution flow
control valve FCV2 and dilution flowmeter FM2.