Options – VICI 450 Dynacalibrator User Manual

Page 10

Advertising
background image

6

calibration output of the cali brator to the inlet port of each of the analyzers. The
calibrator chamber may contain two permeation devices with different trace
gases – one for each analyzer.

The output of the permeation chamber is fed to a mixing tee, where the calibrat-
ed dilution stream is added to set the final concentration. The combined stream
is then fed through another tee to the rear panel overflow vent and span outlet.

Options

The following paragraphs provide general information on each, with more
detailed information found in “Advanced Theory of Operation” on page 27.

Input Configurations

There are six different input plumbing configurations available. The units
may be equipped with or without an internal pump. Units without internal
pumps can be sup plied with one or two gas supply inlet ports (carrier and
dilution). Models with pumps are available with gas stream bypasses for either
the dilution or carrier stream or both. (Units without internal pumps do not
require gas stream bypasses.) Those with internal pumps have a single inlet port
supplying the pump, the outlet of which is split into carrier and dilution streams.
The gas stream bypasses on these units are also downstream from the pump,
permitting the user to connect external scrubbers, driers, separate gas supplies,
etc. in either or both streams. All models are available with nominal maximum
flow capacities from 1 to 20 liters per minute.

Rack Mount Options

The standard Dynacalibrator is packaged in a free-standing case with carrying
handles, for bench use. The units may also be packaged for mounting in a
standard 19" equipment rack.

Special Customer Options

Special options, identified by the “S” at the end of the model number, are not
covered in this basic manual. If your Dynacalibrator is equipped with a special
option, the required additional information will be found on the Special Option
Sheets in the back of this manual. The Special Option Sheets carry the serial
number of the Dynacalibrator to which they apply.

Introduction

Advertising
This manual is related to the following products: