33

Maintenance

Since Dynacalibrators are designed for and generally used in applications that
require continuous service, a planned routine maintenance program is highly
recommended. Routine mainte nance consists of inspection, cleaning, calibration,
and leak checks. The table below lists the recommended checks and mainte-
nance frequencies for continuous and non-continuous use. The recommenda-
tions are provided to help you keep your unit in peak operating condition and
avoid catastrophic failures at inopportune times.

Use the shut-down procedures on page 26 to re move the unit from service and
the turn-on proce dures on page 17 to return it to service.


Maintenance action

By:

Recommended schedule

Continuous use

Periodic use

Instrument inspection and cleaning

User

Quarterly

1000 hours

Leak check

Quarterly

1000 hours

Carrier flow rate verification

Quarterly

1000 hours

Filter inspection and cleaning

Quarterly

1000 hours

Dilution flowmeter calibration

Factory

Annually

4000 hours

Chamber temperature calibration

Annually

4000 hours

Instrument Inspection and Cleaning

The following procedure should be performed quarterly for units in continuous
service or at least after every 1000 hours of service. For the steps which call for
disassembly, see the appropriate paragraph in this section.

1. Remove the unit from service according to the shut down procedures on

page 26. Disconnect all power and plumbing.

2. Remove the top cover of the unit.

3. Snap the black fan screen off the rear and soak it in warm water with a mild

detergent. Do not use a solvent.

4. Remove the four screws, hex nuts, and lockwashers that attach the fan to the

rear panel. Remove the fan filter and soak it with the fan screen in warm water
and a detergent.

5. Blow out the interior of the unit thoroughly with dry, low-velocity air.

Remove any dirt that remains with a soft paint brush.

6. Clean the fan blades with a soft paint brush or a soft cloth dampened in the

water and detergent solution.

7. Clean the exterior surfaces of the unit with a soft cloth dampened in the water

and detergent solution. Do not use an abrasive.

8. Inspect all wiring for any loose connections or broken insulation.