An introduction to colorimetric analysis – LaMotte DCL-20 Colorimeter User Manual

Page 3

Advertising
background image

AN INTRODUCTION TO COLORIMETRIC ANALYSIS

Most test substances in water are colorless and undetectable to the human eye. In order to test for their presence we must

find a way to “see” them. The LaMotte colorimeter can be used to measure any test substance that is itself colored or can be

reacted to produce a color. In fact a simple definition of colorimetry is “the measurement of color” and a colorimetric

method is “any technique used to evaluate an unknown color in reference to known colors”. In a colorimetric chemical test

the intensity of the color from the reaction must be proportional to the concentration of the substance being tested. Some

reactions have limitations or variances inherent to them that may give misleading results. Many such interferences are

discussed with each particular test instruction. In the most basic colorimetric method the reacted test sample is visually

compared to a known color standards. However, accurate and reproducible results are limited by the eyesight of the analyst,

inconsistencies in the light sources, and the fading of color standards.
To avoid these sources of error, a colorimeter can be used to photoelectrically measure the amount of colored light

absorbed by a colored sample in reference to a colorless sample (blank).
Why measure colored light? White light is made up of many different colors or wavelengths of light. A colored sample

typical absorbs only one color or one band of wavelengths from the white light. Not much difference could be measured

between white light before it passes through a colored sample versus after it passes through. The reason for this is that the

one color absorbed by the sample is only a small portion of the total amount of light passing through the sample. However,

if we could select only that one color or band of wavelengths of light which the test sample is most sensitive to, we would

see a large difference between the light before it passes through the sample and after it passes through.
A colorimeter passes a white light beam through an optical filter which transmits only one particular color or band of

wavelengths of light to the photodetector where it is measured. The difference in the amount of colored light transmitted

by a colorless sample (blank) and the amount of colored light transmitted by a colored sample is a measurement of the

amount of colored light absorbed by the sample. In most colorimetric tests the amount of colored light absorbed is directly

proportional to the concentration of the test factor producing the color and the path length through the sample. However,

for some tests the amount of colored light absorbed is inversely proportional to the concentration.
The choice of the correct optical filter and therefore the correct color or wavelength of light is important. It is interesting

to note that the filter that gives the most sensitive calibration for your test factor is the complimentary color of the test

sample. For example, the Nitrate-Nitrogen test produces a pink color proportional to the nitrate concentration in the

sample (the greater the nitrate concentration, the darker the pink color). A green filter is used since a pinkish-red solution

absorbs mostly green light.

REAGENT BLANK

Some tests will provide greater accuracy if a reagent blank is determined, to compensate for any color or turbidity resulting

from the reagents themselves. A reagent blank is performed by running the test procedure on 10 mL of demineralized water.

With the reagent blank in the colorimeter chamber, scan the blank then perform the unknown tests as described.

COLORIMETER TUBES

Colorimeter tubes which have been scratched through excess use should be discarded and replaced with new ones. Dirty

tubes should be cleaned on both the inside and outside. Fingerprints on the exterior of the tubes can cause excessive light

scattering and result in errors. Handle the tubes carefully, making sure the bottom half of the tube is not handled.
LaMotte makes every effort to provide high quality colorimeter tubes. However, wall thicknesses and diameter of tubes may

still vary slightly. This may lead to slight variations in results (e.g. if a tube is turned while in the sample chamber, the

reading will likely change slightly). To eliminate this error put the tubes into the colorimeter chamber with the same

orientation every time.
The tubes that are included with the colorimeter have an index mark to facilitate this.

3

1985-01 •10/01

Advertising
This manual is related to the following products: