SPECIFICATIONS

IN STRU MENT TYPE
Multi-wavelength filter colorimeter _ internal, non-removable filters

READ OUT
3

1

2

inch digit LCD; displays 0–100%T

READ ABLE RES O LU TION
±

1%T

READING STA BIL ITY
±

0.2%T within 5 seconds of turn-on to automatic turn-off

READING IN TER VAL
Approximately 30 seconds with automatic turn-off, resettable

MEA SURE MENT WAVE LENGTHS
1 (420nm), 2 (460nm), 3 (510nm), 4 (530nm), 5 (570nm), 6 (605nm); switch selectable

WAVE LENGTH AC CU RACY
±

1 nanometers

PHO TO MET RIC AC CU RACY
±

0.5%T

SAM PLE CHAM BER
Indexed; accepts 21 mm diameter flat-bottomed test tubes (capped)

SOURCE LAMP
Tungsten filament bulb, 10,000 hour life (est.), spare included, field replaceable

POWER RE QUIRE MENTS
Battery Operation: Field replaceable 1604 type (9V)

Line Operation: 120/220V, 50/60 Hz, 2VA, with optionally-available adapter

DI MEN SIONS
(W x D x H) 190 x 140 90 mm

7

1

2

x 5

1

2

x 3

1

2

inches

WEIGHT
2 lbs.