TROUBLESHOOTING

Problem

Cause

Action

Power on but no display

Batteries
Batteries
Batteries

Insert batteries or remove tab
Verify correct polarity
Replace low or dead batteries

“BAT” message

Batteries

Replace weak batteries

Unstable readings

Insufficientor
depleted
electrolyte

Replace electrolyte, and
membrane cap assembly if
needed

Readings drift down

Insufficient
stirring

Stir sample with meter or use
stirring plate

Slow response

Dirty or
damaged
membrane

Replace electrolyte and
membrane cap assembly

Electrode can not be
calibrated

Depleted
electrolyte
Dirty or
damaged
membrane

Replace electrolyte and
membrane cap assembly
Replace electrolyte and
membrane cap assembly

Electrode can not be
calibrated after replacing
electrolyte and membrane
cap

Dirty probe.
Cathode is
not a shiny
gold color.

Clean cathode with cleaning
cloth or cleaning paper

Sample reading
is frozen

Meter is in
HOLD mode
Meter is
locked

Release HOLD
Remove or replace batteries
and restart. Stored readings
will be lost