TROUBLESHOOTING
Problem
Cause
Action
Power on but no display
Batteries
Batteries
Batteries
Insert batteries or remove tab
Verify correct polarity
Replace low or dead batteries
“BAT” message
Batteries
Replace weak batteries
Unstable readings
Insufficientor
depleted
electrolyte
Replace electrolyte, and
membrane cap assembly if
needed
Readings drift down
Insufficient
stirring
Stir sample with meter or use
stirring plate
Slow response
Dirty or
damaged
membrane
Replace electrolyte and
membrane cap assembly
Electrode can not be
calibrated
Depleted
electrolyte
Dirty or
damaged
membrane
Replace electrolyte and
membrane cap assembly
Replace electrolyte and
membrane cap assembly
Electrode can not be
calibrated after replacing
electrolyte and membrane
cap
Dirty probe.
Cathode is
not a shiny
gold color.
Clean cathode with cleaning
cloth or cleaning paper
Sample reading
is frozen
Meter is in
HOLD mode
Meter is
locked
Release HOLD
Remove or replace batteries
and restart. Stored readings
will be lost