LaMotte TRACER POCKETESTER TDS User Manual

Page 17

Advertising
background image

17

Electrode Cleaning Recommendations
Do not soak the electrode in the solutions for longer than the recommended
length of time. To do so may cause a reference potential shift which will cause
a degradation in performance or failure. When cleaning the probe, do not
scratch or damage the platinized electrode surfaces.

Contaminant

Cleaning Solution

Procedure

Water soluble
substances

Deionized water

Soak or scrub with a soft
brush. Rinse thoroughly with
DI water and dry.

Grease and oil

Warm water and
household detergant

Soak or scrub with a soft
brush, maximun 10 minutes.
Rinse thoroughly with DI
water and dry.

Heavy grease

Alcohol

Soak for a maximum of 5
minutes. Scrub with a soft
brush. Rinse thoroughly with
DI water and dry.

Lime and
hydroxide
coatings and oil

10% Acetic acid

Soak until coating dissolves,
maximum 5 minutes. Rinse
thoroughly with DI water and
dry.

Advertising
This manual is related to the following products: