LaMotte DO 4000 Dissolved Oxygen Meter User Manual

Page 9

Advertising
background image

POLARIZATION
Polarization is the term used to describe when the probe has reached

equilibrium, and is able to produce stable, reliable measurements. For

the probe to be polarized, voltage must continually be applied to the

probe, allowing the oxygen reaction to proceed. When voltage stops, the

reaction no longer continues and the probe is no longer considered

polarized. When either ppm or %SAT is selected on the meter, a voltage

is automatically applied across the electrodes, but the reading is not

accurate until the 15 minute “warm up” time has elapsed. The DO4000

has a constant polarization feature, which allows a small voltage to be

continuously applied to the probe, even when the meter is turned off,

eliminating the 15 minute warm up time. To select the constant

polarization option, set the small button between the connectors to the

OUT position. Since it does use power to keep the probe constantly

polarized, it is recommended this feature not be used if there will be a

span of several hours between measurements.
Do not leave probe polarized for prolonged periods of time

(overnight). Prolonged use will cause membrane fouling and require

more frequent anode cleaning.

9

Advertising