LaMotte WaterLink Spin User Manual

Page 11

Advertising
background image

11

USB CABLE

The USB cable connects the WaterLink Spin to a computer. A USB 2 port is

required.

AC ADAPTER

The AC Adapter supplies power to the WaterLink Spin from an AC outlet.

SET UP

SOFTWARE

Follow the installation instructions included with the software CD.

METER

1. Remove the photometer, AC Adapter and USB cable from the box. Save

this box for shipping purposes if needed in the future.

2. Plug the smaller end of the USB cable (USB mini B connector) into the

meter and the larger end of the USB cable (USB Type A connector) into a

USB 2 port on the computer.

3. Plug the AC adapter into the meter and an AC outlet.

DISKS

1. Each box contains 50 disks for one series, either the chlorine/bromine

series or the biguanide series. Remove a disk from the packaging as

needed.

2. Avoid opening more packages than you need. Disks have a limited

shelf life and should not be exposed to the humidity in the air more than

necessary.

TESTING

1. Select the customer record from the software.
2. Remove a chlorine/bromine or biguanide disk from the package.
3. Use the syringe (1189) to fi ll the disk with the pool or spa water. Add water

until the water column reaches the fi ll line on the disk.

4. Insert the disk into the meter.
5. Cover the disk with the black disk cover (1718).
6. Close the lid.
7. Choose RUN TEST from the computer software program.

SET UP & TESTING