Fcl /+cl – LaMotte ATQ Digital Series Water Softener Demo Kit User Manual

Page 16

Advertising
background image

16

5

/

Fcl /+cl

+ C L

Color

Q

1

+ C L

Color

Q

* Reagent is a potential health hazard. Read MSDS at www.lamotte.com.

1. Press button to go to “FCL” Free Chlorine.

2. Fill a clean tube (0201) to the 5 mL line with water sample.

3. Add

1

DPD 1(6903A) Tablet. Cap. Shake for TEN seconds. Invert

slowly 5 times.

4. Press button to read “FCL” Free Chlorine in ppm. Remove tube.

1. Press button to go to “tCL” Total Chlorine.

2. Remove cap from reacted FCL tube.

3. Add

1

DPD 3(6197A) Tablet. Cap. Shake for TEN seconds. Invert

slowly 5 times.

4. Insert tube into ColorQ.

5. Press button to read “tCL” Total Chlorine in ppm. Remove tube.

1. Press button past “FCL” Free Chlorine to go to “tCL” Total Chlorine.

2. Fill a clean tube (0201)to the 5 mL line with water sample.

3. Add

1

DPD 1(6903A) Tablet. Cap. Shake for TEN seconds. Invert slowly

5 times.

4. Remove cap. Add 1 DPD 3(6197A) Tablet. Cap. Shake for TEN seconds.

Invert slowly 5 times.

5. Insert tube into ColorQ.

6. Press button to read “tCL” Total Chlorine in ppm. Remove tube.

FREE / TOTAL CHLORINE

ColorQ Range: 0.0-10.0 ppm Chlorine

*DPD 1 - Code 6903A-H

*DPD 3 - Code 6197A-H

Free Chlorine and Total Chlorine

Total Chlorine Only

Advertising