Color, User tips, Testing tips – LaMotte ATQ Digital Series Water Softener Demo Kit User Manual

Page 19: Error messages, Maintenance, Battery replacement

Advertising
background image

19

Color

Q

User Tips

Testing Tips

• Read all instructions before testing.

• If not testing all tests in the sequence, follow

blanking procedure and then press button
until desired test factor is shown. Follow test
procedure.

• In one complete inversion, the tube will be

turned cap down and then right side up. The air
bubble will move slowly to the bottom of the
tube and back again to the cap end.

• If small bubbles form after adding reagents,

tap bottom of tube sharply once or twice to
dislodge bubbles. Bubbles will interfere with
the test results.

• Be sure the outside of the tubes are dry before

putting them into the ColorQ. This is best
accomplished by dispensing the water sample
from the sample bottle.

Low pH

Low pH levels can signifi cantly

lower the hardness results.

Error Messages

Er 3

= Insuffi cient light is reaching the detector.

Sample may be too dark, or tube may be in
sideways. Turn off meter and retest.

Er 7

= The blank was not properly set. Turn

off meter and retest blank making sure the black
arrow on the tube is facing forward, toward the
user (not left or right).

SUN

= Extremely bright sunlight is penetrating

the chamber. Turn your back to the sun to shade
the meter from bright sunlight.

Lo

= Result is below the effective measuring

range for the system (e.g. < 4.0 pH for Code
7059 reagent.)

Hi

= Result is above the effective measuring

range for the system (e.g. >9.0 pH for Code 7059
reagent.)

bAt

= Battery is low.

Maintenance

• Clean ColorQ optics with a damp cotton swab.

Avoid abrasive cleaners that can damage
plastic.

• Replace stained or scratched tubes (0201).

Battery Replacement

Use a small Phillips head screwdriver to remove
all 5 screws in the base. Gently pull the battery
holder from the inside and replace TWO AA
style batteries. Replace screws securely but do
not over-tighten screws which can tear the rubber
base.

8/11 62059-01