LaMotte CAUSTIC TEST KIT 7181 User Manual

Caus tic test kit

Advertising
background image

CAUS TIC TEST KIT

CODE 7181

QUANTITY

CONTENTS

CODE

30 mL

*Hydrochloric Acid, 3.6N

*5649WT-G

15 mL

*Phenolphthalein Indicator, 0.5%

*2258-E

30 mL

*Barium Chloride Solution, 10%

*6117-G

1

Pipet, 0.5 mL, plastic

0353

1

Test Tube, 5-10-15-20-25 mL, plastic, w/cap

0715

*WARNING: Reagents marked with an * are considered to be potential health
hazards. To view or print a Material Safety Data Sheet (MSDS) for these reagents go to
www.lamotte.com. To obtain a printed copy, contact LaMotte by e-mail, phone or fax.
To order individual reagents or test kit components, use the specified code number.

NOTE: The procedures below eliminate non-hydroxyl alkalinity
contributors. If a total alkalinity determination is preferred, skip the
addition of *Barium Chloride Solution 10% (6117) in Step 2 or Step 3.
Record the result as total alkalinity as NaOH.

PRO CE DURE

1 DROP = 0.1%
1.

Fill test tube (0715) to 5 mL line with sample.

2.

Add 10 drops of *Barium Chloride Solution, 10% (6117). Swirl to
mix. A white precipitate will form if carbonates are present.

3.

Add 1 drop of *Phenolphthalein Indicator, 0.5% (2258). Swirl to
mix. Solution will turn pink.

4.

While gently swirling tube, add *Hydrochloric Acid, 3.6N
(5649WT) one drop at a time until pink color disappears. Count the
number of drops added. Hold dropper bottle vertically.

5.

Multiply the number of drops used in Step 4 by 0.1. Record as
Percent Caustic as NaOH.

Advertising