Lamotte com pany – LaMotte CAUSTIC TEST KIT 7181 User Manual

Page 2

Advertising
background image

1 DROP = 1%
1.

Use the 0.5 mL pipet (0353) to add 0.5 mL of sample to test tube
(0715).

2.

Dilute to 5 mL line with tap water.

3.

Add 10 drops of *Barium Chloride Solution, 10% (6117). Swirl to
mix. A white precipitate will form if carbonates are present.

4.

Add 1 drop of *Phenolphthalein Indicator, 0.5% (2258). Swirl to
mix. Solution will turn pink.

5.

While gently swirling tube, add *Hydrochloric Acid, 3.6N
(5649WT) one drop at a time until pink color disappears. Count the
number of drops added. Hold dropper bottle vertically.

6.

Multiply the number of drops used in Step 5 by 1. Record as
Percent Caustic as NaOH.

CON VER SION
To report the result as Percent Caustic as KOH, multiply the Percent
Caustic as NaOH by 1.4

Percent Caustic as NaOH X 1.4 = Percent Causic as KOH

For example:

0.2% as NaOH X 1.4 = 0.28% as KOH

LaMOTTE COM PANY

Helping Peo ple Solve An a lyt i cal Challenges

®

PO Box 329 • Chestertown • Mary land • 21620 • USA

800-344-3100 • 410-778-3100 (Out side U.S.A.) • Fax 410-778-6394

Visit us on the web at www.lamotte.com

1.11