Pro ce dure
Low Range (0.00 – 0.20 ppm NO

3

_N)

1.

Firmly tap the capped Enzyme Vial (8136) three times on a hard surface.
This will ensure that reagents are at the bottom of the vial.

2.

Use the 1 mL pipet (0354) to add 2 mL of Nitrate Buffer (8139).

3.

Use a plain pipet (0352) to add sample water until the vial is filled to the
5 mL line.

4.

Cap the vial and mix by inverting gently 3 times. Repeat the mixing
procedure at 5 minutes, 10 minutes and 15 minutes. Total reaction time
is 15 minutes.

5.

Add 5 drops Hydrochloric Acid, 1N (6130). Cap and mix.

6.

Add 1 Nitrate Color Tablet (8140A). Cap the vial and mix by gently
inverting until the tablet disintegrates.

7.

Wait 5 minutes.

8.

Remove the cap. Place the vial on the white area of the Nitrate-N Low
Range Color Chart (8137LR-CC). Looking down through the tube,
compare the color of the solution to the color chart.

9.

Record the result as ppm Nitrate-N (NO

3

_N). To convert the result to

Nitrate (NO

3

) multiply by 4.4.

NOTE:

Salinity will reduce the enzyme effectiveness and result in a decrease of

color development.
NOTE:

For the most accurate results use laboratory-grade pipets for measuring

sample water and reagents.