LaMotte Nitrate-Nitrogen Low Range Enzyme Kit User Manual

Page 5

Advertising
background image

Pro ce dure
High Range (0.2 – 1.0 ppm NO

3

_N)

1.

Firmly tap the capped Enzyme Vial (8136) three times on a hard
surface. This will ensure that reagents are at the bottom of the vial.

2.

Use the 1 mL pipet (0354) to add 2 mL of Nitrate Buffer (8139).

3.

Use the 0.5 mL pipet (0353) to add 0.5 mL of the sample water to
the Enzyme vial.

4.

Use a plain pipet (0352) to add Deionized Water (5115) until the
vial is filled to the 5 mL line.

5.

Cap the vial. Mix by inverting gently 3 times. Repeat the mixing
procedure at 5 minutes, 10 minutes and 15 minutes.The total
reaction time is 15 minutes.

6.

Add 5 drops Hydrochloric Acid, 1N (6130). Cap and mix.

7.

Add 1 Nitrate Color Tablet (8140A). Cap the vial and mix by gently
inverting until the tablet disintegrates.

8.

Wait 5 minutes.

9.

Remove the cap. Place the vial on the white area of the Nitrate-N
High Range Color Chart (8137HR-CC). Looking down through the
tube, compare the color of the solution to the color chart.

10.

Record the result as ppm Nitrate-N (NO

3

_N). To convert the result

to Nitrate (NO

3

) multiply by 4.4.

NOTE:

Salinity will reduce the enzyme effectiveness and result in a decrease

of color development.
NOTE:

For the most ac cu rate re sults use lab o ra tory-grade pipets for

mea sur ing sam ple wa ter and re agents.

Advertising