Amico Light-Duty Monitor Arm User Manual

Page 16

Advertising
background image

16

Amico Accessories Inc.

SECTION 6: Maintenance and Troubleshooting

Cleaning

WARNING

:

The cleaning chemicals and methods below are not meant for controlling any infections. It shall

be the responsibility of the hospital or the hospital’s infection control officer to sanitize the equipment.

WARNING

:

Please do not spray any chemical directly onto the arm. Apply onto a soft cloth and wipe clean

to prevent chemicals getting into the internal components of the arm.

The mounting assembly may be cleaned with most mild, non-abrasive solutions commonly used in the hospital
environment (e.g. diluted bleach, ammonia, or alcohol solutions). The surface finish will be permanently damaged by
strong chemicals and solvents such as acetone and trichloroethylene. Steel wool or other abrasive material should never
be used.

Damage caused by the use of unapproved substances or processes will not be warranted. It is recommended that you
test any cleaning solution on a small area of the LMA that is not visible to verify compatibility.

Never submerge the LMA and do not allow liquids to enter it. Wipe any cleaning agents off the LMA immediately using
a water-dampened cloth. Dry the arm thoroughly after cleaning.

No Acetone

Acetone is a colorless,
mobile, flammable liquid

Acetone

No Trichloroethylene

Trichloroethylene is a chlorinated
hydrocarbon commonly used as
an industrial solvent.

Trichloroethylene

Advertising