Amico NFPA Triplex RVL Modular Stack Mount User Manual

Page 19

Advertising
background image

www.amico.com

3.

Panel designed with selectable options to fully match applications

a.

All system settings shall be user adjustable and accessible with the system in operation and

the control panel door closed (password protected).

b.

Adjustments can be made to pressure settings to match the customers’ requirements.

c.

All alarms to require manual reset.

Specifier: select the section below reflecting the preferred technology:

Standard alarms shall open on failure with local audible and visual alarms with dry contacts for the following
conditions:

4.

OIL-LESS CLAW PUMP

a.

Lag pump in use

b.

Main transformer failure

c.

Motor

overload

d.

High discharge air temperature

e.

High inlet vacuum

4.

LUBRICATED ROTARY VANE PUMP

a.

Lag pump in use

b.

Main transformer failure

c.

Motor

overload

d.

High discharge air temperature

e.

Low oil level

4.

DRY ROTARY VANE PUMP

a.

Lag pump in use

b.

Main transformer failure

c.

Motor

overload

d.

High discharge air temperature

4.

WATER-SEALED LIQUID RING VACUUM PUMP

a.

Lag pump in use

b.

Main transformer failure

c.

Motor

overload

d.

High water temperature

17