Amico NFPA Triplex RVL Modular Stack Mount User Manual

Page 8

Advertising
background image

Amico Source Equipment

7.3.8 Specification

1.

Select the sections appropriate to the technology and system architecture desired.

2.

Write into the specification any exceptional requirements (soft starters, etc.) that are necessary.

3.

Schedule on the drawings the vacuum plant selected.

Schedule at least:

a.

The capacity per pump (per NFPA) and total system capacity.

b.

Horsepower per pump or Kilowatt.

c.

Voltage, Hz, and phase desired.

7.3.9 Technology Comparisons

Amico Source offers several technologies for medical vacuums, each of which has its own advantages and
drawbacks. The page summarizes the features of these technologies as an aid in the selection of the correct
technology for your specific application. Please refer to Figure 3 for table.

6

Advertising