Initial power-up – Amico Automatic Cryogenic Liquid Solenoid Manifold User Manual

Page 19

Advertising
background image

www.amico.com

19

Initial Power-Up

Release the two latches on either side of the manifold control panel and remove the cover.

Verify the following:
• Both master valves located on both header bars are open
• All four line regulator isolation valves are open (handles in horizontal position)
• Power supply has been connected
• Both Red “Empty” indicators on the front of the manifold are illuminated
• If connected to a master alarm panel, “Secondary Supply” alarm is activated

Close 3/4” source shut-off valve.

Slowly open one tank on the right side of the manifold and press the “Push to Reset” button between
the bank LED indicators.

Verify the following:
• Right bank Red “Empty” LED goes out
• Right bank Green “In Use” LED illuminates
• Right bank cylinder contents gauge reads cylinder pressure

Slowly open one tank on the left side of the manifold and press the “Push to reset” button between the bank indicators.

Verify the following:
• Left bank Red “Empty” LED goes out
• Left bank Yellow “Ready” LED illuminates
• Left bank cylinder contents gauge reads cylinder pressure
• If connected to a master alarm panel, “secondary supply” alarm is not activated

Close right bank cylinder. Slightly depress bleed valve located on the side of the regulator. Verify the following:
• Right bank cylinder contents gauge drops slowly
• As the right cylinder contents gauge is nearly depleted, the manifold changes over to the left bank
• After change-over, the right bank Green “In Use” LED goes out and the Red “Empty” LED illuminates
• After change-over, the left bank Yellow “Ready” LED goes out and the Green “In Use” LED illuminates

Verify the “line pressure” gauge reading is acceptable.

Slowly open one cylinder on the right side of the manifold and press the “Push to Reset” button.

1.

2.

3.

4.

5.

6.

7.

8.

9.


10.