Latch valve assembly oxequip/medstar (compatible) – Amico Retractable Ceiling Column Outlet User Manual

Page 12

Advertising
background image

Latch Valve Assembly Oxequip/MEDSTAR (Compatible)

1. Unscrew the two retaining screws (12) until the Latch Valve Assembly (11) can be removed from the outlet.

2.

Locking Mechanism Repair

Unscrewing the first screw (7) and remove item (6). Then unscrew the second set of screws (7), this will give you

access to items (8), (9) and (10). Inspect all three items for wear or damage, replace if needed.

3.

Outlet Gas Connector Repair.

While holding the round retaining ring (19) in place, remove the two screws (20). Remove the retaining ring (19)

and gas connector (18). Then remove the Body Seal (15), the Poppet (16) and the Gas or Vacuum Spring (17).

Inspect for wear or damage, replace components if needed.

Re-install the Gas or Vacuum Spring (17), the Poppet (16) and the Body Seal (15) into the gas connector (18).

Re-install gas connector (18) and retaining plate (19), secure with two screws (20). DO NOT OVERTIGHTEN.

4. Re-install the Latch Valve Assembly into the outlet. Coat the connector (18) with a thin coat of Oxygen compatible

silicone lubricant to aid insertion. Tighten down the retaining screws (12), DO NOT over tighten, as this could

damage the Latch Valve.

5. Connect a gas specific adapter into the outlet. The connection should be smooth and the adapter should lock

and remain in place allowing gas to flow. If not, replace the entire Latch Valve Assembly (11).