Installation and dimensions, Inch [mm – Amico Retractable Ceiling Column Outlet User Manual
Page 7
Ceiling Column Outlets
DISS, Ohmeda compatible, Chemetron compatible, Puritan-Bennett and Oxequip/MedStar compatible
On all outlets (wall and console) the connecting pipe can swivel a full 360 deg. for easy connection to the gas pipe.
When soldering the pipe connection take care not to heat up the body, since that could damage the secondary check
valve.
It is very important to keep the protective cover (wall outlets only) in place during construction, to ensure that no debris
or dirt get into the outlet.
When installing the “Latch-valve assembly”, remove the protective cover and inspect for dirt or debris in the outlet body.
Clean out the contaminants, if required.
The “Latch-valve assembly” should slide in smoothly into the “Rough-In assembly”, if not, check that there is no damages to
the indexing pin arrangements. If the indexing pins are bent or damaged, replace the “Latch-valve assembly”.
DO NOT PRESSURE TEST THE PIPELINE OVER 100 PSI [690 kPa] IF THE “LATCH-VALVE ASSEMBLY” HAS BEEN
INSTALLED.
All “Rough-in assemblies” can be pressure tested up to a maximum of 200psi without the “Latch-valve assembly” attached
to it.
All “Latch-valve assemblies” with the exception of Nitrogen DISS have a Maximum pressure rating of 100 PSI [690 kPa]. The
Nitrogen D I S S has a maximum pressure rating of 200 psi [1,380 kPa].
After installation, perform the test on page 1.
1/2" [12.7 mm] O.D. (3/8” nominal) type "K"
copper pipe, rotates 360 deg.
for entry from any angle.
9-1/4"
[235]
SIDE VIEW
2-7/16"
[62]
SIDE VIEW
FRONT
ASSEMBLY
3-3/4"
[95]
FRONT VIEW
FRONT
ASSEMBLY
2-7/16"
[62]
3-9/32"
[83]
1-13/16"
[46]
3-9/16"
[91]
FRONT VIEW
BACK
ASSEMBLY
OXYGEN
OXY
OXY
inch
[mm]
Installation and Dimensions
www.amico.com
7