Safety and regulatory compliance – Bio-Rad T100™ Thermal Cycler User Manual

Page 7

Advertising
background image

T100 Thermal Cycler Manual

v

Safety and Regulatory Compliance

This instrument has been tested and found to be in compliance with all applicable
requirements of the following safety and electromagnetic standards (Table

5).

S

AFETY

C

OMPLIANCE

This instrument has been tested and found to be in compliance with all applicable
requirements of the following safety and electromagnetic standards:

• IEC 61010-1:2001 Safety requirements for electrical equipment for measurement,

control, and laboratory use —Part 1: General requirements

• IEC 61010-2-010:2003 Safety requirements for electrical equipment for

measurement, control, and laboratory use —Part 2-010: Particular requirements
for laboratory equipment for the heating of material

• CAN/CSA-C22.2 NO. 61010-1-04 Safety Requirements for Electrical Equipment

for Measurement, Control, and Laboratory Use - Part 1: General Requirements

• CAN/CSA-C22.2 NO. 61010-2-010-04 Safety Requirements for Electrical

Equipment for Measurement, Control, and Laboratory Use - Part 2-010: Particular
Requirements for Laboratory Equipment for the Heating of Materials

• EN 61010-1:2001 Safety requirements for electrical equipment for measurement,

control, and laboratory use —Part 1: General requirements

Never remove the outer case of a thermal cycler base. This may cause you to
receive an electrical shock.
This thermal cycler uses neutral fusing, which means that live power could still be
exposed inside the instrument even when the fuse is blown or removed.

Warning about risk of burning.
A thermal cycler generates enough heat to cause serious burns. Wear safety
goggles or other eye protection at all times during operation. Always allow the
sample block to return to idle temperature before opening the lid and removing
samples. Always allow maximum clearance to avoid accidental skin burns.

Warning about risk of explosion.
The sample blocks can become hot enough during the course of normal operation
to cause liquids to boil and explode.

Table

5. Safe use specifications.

Usage Aspect

Conditions for Safe Use

Rated input power

100–240 Vac, 50–60 Hz, 700 W Max

Overvoltage category

II

Replacement fuses

10 A, 250 V, 5 x 20 mm, fast blow (qty 2)

Environment

Indoor use only

Temperature

15–31°C

Relative humidity

Up to 80% (noncondensing)

Altitude

Up to 2,000 meters above sea level

Pollution degree

2

Table

4. Instrument safety warning labels. (continued)