Silver Stain Troubleshooting

Problem

Cause

Solution

7. Dark background

a. Low quality reagents

a. Contaminants will

or contaminated

catalyze non-specific

water

formation of metallic
silver. Use high purity
reagents and deionized
water

b. Over development

b. Monitor the development

of the gel

step carefully and stop
quickly by draining
through both drain ports
(stopcocks). Then
quickly add the stop
solution through the
reagent access door of
the lid. The drain/fill
cycle should be no
longer than 3 minutes

c. Overloaded protein

c. Adjust the protein sample

sample

load so it is appropriate
for this detection
method

d. Incorrect preparation

d. Refer to the silver stain

of silver stain

instruction manual,

solutions

paying special attention
to the image developer
concentrate amounts

8. Poor sensitivity

a. Inactive reagents or

a. Use the silver stain kit

inaccurate dilution or

components within their

incorrect incubation

recommended shelf life

times

Confirm solution
preparation and
incubation times are
correct. Pay special
attention to the amount
of image developer
concentrate

b. Low protein sample

b. Adjust the protein sample

load

load so it is appropriate
for this detection
method

Sypro Ruby Troubleshooting

Problem

Cause

Solution

9. Weak or no

a. Residual Coomassie

a. Clean the Dodeca stainer

fluorescent signal

stain in the system

thoroughly before

inhibits the fluorescent

switching from

signal

Coomassie based stain
to fluorescent stains

b. Low protein sample

b. Adjust the protein sample

load

load so it is appropriate
for this detection method