Bio-Rad PowerPac™ HC High-Current Power Supply User Manual

Page 15

Advertising
background image

If a power failure occurs during a run, there are several possible scenarios which
are described as follows:

Message Displayed

Explanation

RUN ABORTED

PFd was OFF.

AFTER POWER FAIL

The run, either timed or continuous, was aborted after a power
failure occurred.

*PRESS STOP KEY

After pressing the stop key to reset, the original run parameters

TO RESET

are no longer available.

RUN STOPPED

PFd was ON when a power failure occurred and the run was

AFTER POWER FAIL

resumed automatically.
The run, either timed or continuous, has been terminated by
pressing the stop key.

(Message alternates

After pressing the stop key to reset, the original run parameters

with final run parameters) are available for re-run.

Note: PFd defaults automatically to OFF after pressing the stop
key.

RUN COMPLETED

PFd was ON when a power failure occurred and the run was

AFTER POWER FAIL

resumed automatically.
The run was timed and automatically terminated when the set

(Message alternates

run time was completed.

final run parameters)

After pressing the stop key to reset, the original run parameters
are available for re-run.
Note: PFd defaults automatically to OFF when the run is
completed.

4.2.3 User Serviceable Error Messages
The following user serviceable messages may appear, interrupting a run. Refer to
this section for their cause and solution. Follow instructions on the screen to clear
the error message and resume operation.

Problem/error
message

Cause Solution

E01 NO LOAD

No load detected.

Verify all electrical connections.

DETECTED

Electrophoresis cell

Verify buffer levels are appropriate.

not connected to the

CELL & CABLES

power supply.

Verify the electrophoresis application's

* PRESS STOP KEY

power requirements match PowerPac

The current load is

HC output range.

less than 0.01 A.

E02 SHORT CIRCUIT Short circuit detected

Insure all electrical connections

(over current, load

are making good contact and the

CELL & CABLES

current is greater

cables and wire electrodes are in

* POWER OFF/ON

than 3.0 A).

good shape.

Accidental shorting of

Verify buffer levels are appropriate.

output leads.

Shorting due to wrong
connections.

E04 REGULATING

Regulation Error.

Ensure the electrophoresis application

BELOW 5V

Current drawn by the

settings and power requirements

electrophoresis system

match PowerPac HC output range

METHOD VALUES exceeds the set current

* PRESS STOP KEY limit of the unit, at less

than 5 V.

9

Advertising