Problem

Cause Solution

E8 error code

Regulation error

Turn power supply off, then on to

reset.
displayed

If problem persists, contact Bio-Rad
Technical Resources

E9 error code

Change in Load

Check and correct any potential

displayed

Resistance.

resistance problem then,
• press run/pause key to resume run

or,

The PowerPac

• press any key, other than the

run/pause

Basic detects drastic

key, to clear the error code.

changes in resistance
which may indicate
failure of the cell’s power
leads or a loose output
connection.

Note: Certain applications
exhibit fluctuations in
resistance that can
trigger the change in
resistance error code.
If this is the case, the
change in resistance
feature can be deactivated
to allow uninterrupted
completion of a run.
(see Section 3, Step 8)

Caution: Deactivating this
safety feature increases the
possibility that a failure of the
power leads or loose output
connection will not be
detected.

E10 error code

Unacceptable value(s)

Clear the code by pressing any key

displayed

entered

other than the run/pause key.
Then, enter values within range of
PowerPac Basic and press the
run/pause key

E12 error code

Internal Over Current

Possible power supply malfunction.

displayed
E13 error code

Internal Short Circuit

Check for and correct problems such

displayed

as dirty contacts, frayed wires,

E14 error code

Internal Over Voltage

excessive buffer concentration. Then

displayed

press any key other than the run/pause

E15 error code

Internal Short Circuit

key to clear the code.

displayed
E16 error code

Hardware failure

Contact Bio-Rad Technical Resources

displayed
E17 error code

Hardware failure

displayed
E98 error code

Internal system error

displayed
E99 error code

Internal system error

displayed