Section 2
Specifications

Cooling Module

Chassis

aluminum

Dimensions

42 cm long x 23 cm wide x 24 cm high

Weight

14 kg

Compressor

1/12 HP

Refrigerant

CFC 12

Heat exchanger

stainless steel

Operating range

5 °C–25 °C

Cooling rate (pre-cool)

0.75 °C/minute from ambient to 14 °C

Cooling capacity (input power)

75 watts at 14 °C

Input voltage range

90-132 VAC/47-63 Hz/4 amps
198-264 VAC/47-63 Hz/2 amps

Fuses

6.3 amp Slow-Blow (100/120 V) or
3.0 amp Slow-Blow (220/240 V)

Note: This equipment has been tested and found to comply with the limits for a Class A
digital device, pursuant to Part 15 of the FCC rules. These limits provide reasonable protection
against harmful interference when the equipment is operated in a commercial environment.
This equipment generates, uses, and can radiate radio frequency energy and, if not installed
and used in accordance with the instruction manual, may cause harmful interference to radio
communications. Operation of this equipment in a residential area is likely to cause harmful
interference in which case the user will be required to correct the interference at his own
expense.

Section 3
Safety Information

The Cooling Module is operated in conjunction with an electrophoresis apparatus

connected to a high voltage power supply. Although those components of the Cooling Module
which are in contact with the electrically charged buffer are electrically isolated, certain
precautions should be taken to eliminate electric shock hazard.

1) The power supply should be off when connecting or disconnecting tubing.

2) The tubing must be firmly seated against the rubber grommets on the “FLOW IN” and

“FLOW OUT” ports to prevent accidental contact with the metal tubing of the heat
exchanger.

3) The safety interlock of the electrophoresis chamber lid should never be bypassed to

permit operation of the instrument without the lid.

!

2