Bio-Rad recommends that the buffer in the system be pre-cooled to the desired run

temperature. The Cooling Module can quickly pre-cool the system, at an approximate rate
of 0.75 °C/minute in the absence of input power to the recommended operating temperature
of 14 °C (to pre-cool from ambient temperature to 14 °C only takes 10–15 minutes), and is
designed to maintain set temperature during the run. If the buffer is not pre-cooled, the
presence of input power (from the electrophoresis power supply) will greatly increase the
time required to bring the system to the set temperature.

Note: The 0.75 °C/minute pre-cooling rate decreases for set temperatures lower than
14 °C.

Section 6
Troubleshooting Guide

Problem

Solution

No power

Check fuse in main power switch on back panel of unit (see
Section 7)

Fuse in main power switch OK; contact Bio-Rad instrument
service

Power on, compressor

Reset timer

not running

Temperature set too high; lower setting

Still in 100 second delay period

No or low buffer flow

Check tubing for kinks

Pump flow rate too low

Frozen buffer in heat exchanger; turn pump on; raise set tem-
perature above ambient until buffer flow resumes, then lower
set temperature to desired setting.

Insufficient cooling

Pre-cool buffer to set temperature prior to beginning gel run

Buffer concentration too high

Buffer flow rate too high

Electrophoresis or buffer flow conditions outside design range
limits

Unusual noise

Frozen buffer in heat exchanger; in variable speed pump see

or vibration

above

Air leak from loose tubing connection

Small pool of liquid

Frost may form on heat exchanger under certain operating

under Cooling Module

conditions resulting in small amounts of water following
Cooling Module shutdown.