Bio-Rad CHEF-DR® III Variable Angle System User Manual
Page 29

26
5. Hansenula wingei
Size Range:
1–3.1 mb
Agarose:
0.8% Molecular Biology Certified
Buffer:
1.0x TAE
Temperature:
14 °C
Switch Time:
500 seconds
Run Time:
48 hours
Angle:
106°
Voltage Gradient:
3 V/cm
6. Schizosaccharomyces pombe
Size Range:
3.5–5.7 mb
Agarose:
0.8% Chromosomal Grade
Buffer:
1.0x TAE
Temperature:
14 °C
Switch Time:
1,800 seconds
Run Time:
72 hours
Angle:
106°
Voltage Gradient:
2 V/cm
7. Angle Ramp: S. pombe
Size Range:
3.5–5.7 mb
Agarose:
0.8% Chromosomal Grade
Buffer:
1.0x TAE
Temperature:
14 °C
Block 1
Block 2
Block 3
Switch Time:
1,200 sec.
1,500 sec. 1,800 sec.
Run Time:
24 hours
24 hours
24 hours
Angle:
96°
100°
106°
Voltage Gradient:
2 V/cm
2 V/cm
2 V/cm
8. Voltage Ramp: S. pombe
Size Range:
3.5–5.7 mb
Agarose:
0.8% Chromosomal Grade
Buffer:
1.0x TAE
Temperature:
14 °C
Block 1
Block 2
Block 3
Switch Time:
1,800 sec. 1,500 sec. 1,200 sec.
Run Time:
24 hours
24 hours
24 hours
Angle:
106°
100°
96°
Voltage Gradient:
1.5 V/cm
2 V/cm
2.5 V/cm