Table vii: hardware iq test – Bio-Rad Experion DNA Analysis Kits User Manual

Page 27

Advertising
background image

Technical Support: 1-800-424-6723 • www.bio-rad.com

19

Installation Qualification

Experion Electrophoresis Station – Serial Number______

Table VII: Hardware IQ Test

Item

Description

Performed by/date

1

Turn on the Experion Station. Boot up the Experion software. Select
Hardware IQ through the Validation Menu. Select Electronic and
Communication Tests.

2

Select Start. All tests must be executed. It is not possible to deselect a
test. If a failure results, repeat test. If a second failure occurs, call Bio-Rad
Technical Support.

3

Select Next to view details. Select Print to print the report on the next page.
Security Edition only: If further IQ/OQ tests will be performed, print all test
results together through Report IQ/OQ Results.

4

Append print report to IQ/OQ records binder.

Comments:_________________________________________________________________________
Performed by:_________________________________ Date:_______________________________
Reviewed by:__________________________________ Date:_______________________________