Bio-Rad Gel Doc™ XR+ System User Manual
Page 23
Advertising

Page 23 of 42
1. Make sure that all lights in the Universal Hood II are off.
2. Make sure that the Lens Cap is on the lens. This is critical!
3. Power ON the Camera and the Universal Hood II.
4. Wait for 30 minutes for the system to warm up and stabilize
5. Double click on Quantity One
software icon on the desktop. From “File Menu” select
Gel Doc
XR OR ChemiDoc
XRS.
TDS Quantity One
Window
Select Gel Doc
XR OR ChemiDoc XRS from
File Menu
5. The following window will open with a notification that the system will be acquiring Dark
Reference images.
Advertising
This manual is related to the following products: