Section 8 Maintenance and Part Replacement

This section covers the replacement of parts in the Universal Hood II.

8.1 Epi-illumination lamp replacement

The lamps are located behind the two panels on each side (left and right) of the internal side of
the hood.

•

Turn system power off.

•

Remove the power cord.

•

Open the enclosure door.

•

Locate the Epi lamp housing.

•

Locate the socket head screw that hold each lamp housing to the internal side of the hood.

•

Remove this screw.

•

The cover screen with a plastic piece will come loose. Pull on the cover to remove it from

the box and the lamp will become visible.

•

To remove the lamp, hold it from the receptacle and then pull it from the plastic Holder.

•

Insert the new lamp (5 or 11 W) into the lamp holder and then push it into the receptacle.

•

Reassemble the cover.

Epi-White Light 5 W assembly (left side)

8.2 Fuse replacement

Always unplug the instrument before changing or checking the fuses.



Power Entry Module

Power Cable plug-in

ON/OFF Main Switch

Fuse holder