Error messages – Bio-Rad Image Lab™ Software User Manual
Page 179
Advertising

Error Messages
User Guide
| 179
Error Messages
Sample tray not detected
The sample tray has a magnet that senses when the sample tray is inserted. If you
attempt to image and this message appears, the sample tray might not be pushed
in all the way. Press the tray in until the magnet grabs the tray.
Light tray not detected
This error message appears if the lamp assembly is not fully seated into the imager.
Reseat the lamp assembly as described on the previous page.
Door was opened during imaging
To prevent UV radiation exposure, the UV-B lamps do not turn on unless the front
sample tray door is fully closed. This message also appears when the instrument
cover has been removed.