11

2. Select OK and follow the software prompts for step-by-step

instructions for CAL1 and CAL2 calibration.

Note:

In Bio-Plex Manager version 6.1 and higher, startup, warm up,

and calibration can be performed together by selecting the “Start up and
calibrate” icon.

3. Prepare Wash Method

Bio-Plex Pro

assays are compatible with both magnetic separation and

vacuum filtration methods. However, for best results, we recommend
performing the assays in a flat bottom plate with magnetic separation.

Setting Up the Bio-Plex Pro or Bio-Plex Pro II

Wash Station

The wash station should be primed before use. For more information, refer
to the Bio-Plex Pro Wash Stations Quick Guide (bulletin #5826).

1. Install the appropriate plate carrier on the wash station.
2. Use the prime procedure to prime channel 1 with 1x assay buffer.

Table 3. Summary of compatible wash stations and plate types.
Wash Method

Wash Station

Assay Plate

Magnetic separation

Bio-Plex Pro

Flat bottom plate

Bio-Plex Pro II (use MAG programs)

Bio-Plex

®

handheld magnetic washer

Vacuum filtration

Bio-Plex Pro II (use VAC programs)

Filter plate

Vacuum manifold (manual)