Bio-Rad Human Metabolic and Hormone Assays User Manual

Page 6

Advertising
background image

4

Fig. 1. Bio-Plex sandwich immunoassay.

Data Acquisition and Analysis
Data from the reactions are acquired using a Bio-Plex system or similar
Luminex-based reader. When a multiplex assay suspension is drawn into
the Bio-Plex 200 reader, for example, a red (635 nm) laser illuminates the
fluorescent dyes within each bead to provide bead classification and thus
assay identification. At the same time, a green (532 nm) laser excites PE
to generate a reporter signal, which is detected by a photomultiplier tube
(PMT). A high-speed digital processor manages data output and
Bio-Plex Manager

software presents data as median fluorescence

intensity (MFI) as well as concentration. The concentration of analyte
bound to each bead is proportional to the MFI of reporter signal.

Biomarker

of Interest

Magnetic Bead

Capture

Antibody

Biotinylated

Detection

Antibody

Phycoerythrin

Fluorescent

Reporter

Streptavidin

Advertising