Bio-Rad Bio-Rex® 5 Anion Exchange Resins User Manual

Page 8

Advertising
background image

3.

Neutral fraction.

a.

Collect in a clean vial (scintillation vials work
well) and use a syringe to hasten flow through
the column.

b.

Add the sample to column, drain through.

c.

Wash with water (1 ml, 4 ml, 4 ml).

d.

Cap and mix well. Store at 4 °C.

4.

Acid fraction.

a.

Change to a fresh vial.

b.

Carefully add 1 ml of 10% sulfuric acid.

c.

Wash with water (1 ml, 4 ml, 4 ml).

d.

Cap and mix well.

Technical Information

If you have any questions or need additional

information, contact your local Bio-Rad representative.

13

Since some of these compounds co-elute during HPLC,
this simple fractionation technique was developed.

Protocol

1.

In a test tube, carefully pipet:

1.00 ml wine or grape must

0.100 ml 5% formic acid

0.100 ml 5% propylene glycol

0.2 ml 5 N ammonium hydroxide

Mix well. Insure that the pH is between 8 and 9;
adjust if necessary. Keep closed.

2.

Prepare the column as follows (do not let the column
dry out):

a.

Slurry 1 g Bio-Rex 5 100-200 mesh resin (Cl

-

form) in 3 ml water.

b.

Pour slurry into a polypropylene column.

c.

Using a syringe, expel excess liquid, then wash
bed with 5 ml water.

12

LIT207B 9/2/98 10:03 AM Page 12

Advertising
This manual is related to the following products: