1.6 mA meter excitation (S2P, S2V)

NOTES: 1) Maximum excitation for “Ratiometric” measurement is 10 Vdc.

2) Remove the “S2T” jumper when you use external sensor excitation.

6.4.1 Jumper Settings for Sensor Excitation –ac Powered Unit

Your ac-powered meter is capable of supplying either 1.5 to 11 Vdc or 24 Vdc sensor excitation. (refer to Figure 6-
2).

• For 1.5 to 11 Vdc, install S4A and C jumpers, then adjust the potentiometer (R39) for proper voltage.

• For 24 Vdc excitation, remove S4A and C jumpers and install S4B.

Figure 6-2. ac-Powered Main Board Jumper positions S3 and S4

S4 On ac-powered units, main board
S4A & S4C

Installed

For 1.5 to 11 Vdc excitation

S4B

Installed

For 24 Vdc excitation

S3 On ac-powered units, main board
S3A

Installed

Unlocks “MENU” button for programming

S3B

Omit

S3C

Installed

Unlocks lockout menu


NOTE: S3B (the middle jumper) should NOT be installed. This jumper is only used when recalibrating the meter
(e.g. an annual, careful performance by the calibration lab). When this jumper is installed, calibration coefficients
can be changed via digital communications.

6.4.2 Jumper Settings for Sensor Excitation – dc Powered Unit

Your dc-powered meter is capable of supplying either 1.25 to 12 Vdc or 24 Vdc sensor excitation. (Refer to Figure
6.3)

• For 1.25 to 12 Vdc excitation, install SA and S4A jumpers, then adjust the potentiometer (R34) for proper

voltage.

• For 24 Vdc, remove the S4A jumper s and install S1B.

CF 67

17

M1291/N/0403 11279ML-02 Rev. A