Trouble Shooting Guide
OPERATOR’S MANUAL
CenSlide 2000 Urinalysis System
Page 5 - 1
5.0 Trouble Shooting
OBSERVATIONS
ACTIONS
The CenSlide 2000 Centrifuge
Check that the power transformer
does not appear to have power.
is securely plugged into both the AC wall
outlet and the back of the centrifuge.
Try a different AC outlet to verify that
AC current is being supplied to the
power supply module.
Make sure the lid is securely closed and
locked in place.
Reset the CenSlide 2000 Centrifuge by
disconnecting and reconnecting the power
supply cord's plug to the centrifuge.
The CenSlide 2000
Balance appropriately by placing equally
Centrifuge appears to
filled CenSlide tubes opposite each
vibrate when operating.
other.
It does not appear that
Seat CenSlide tubes firmly and securely
the flicking mechanism
within the CenSlide 2000 Centrifuge
is functioning.
rotor.
Check that you are using the correct
power supply module. Use only the power
supply provided with the instrument.
Clean and dry the flicker button with alcohol
and a tissue or swab.