Operating the scale, Calibration/adjustment, Basic weighing function – CITIZEN CARAT BALANCES CT103 User Manual

Page 3

Advertising
background image

3

Calibration/Adjustment*

Always calibrate/adjust the scale after setting it up in a new

location.

Available Features

Calibration/adjustment can only be performed when

– there is no load on the scale,

– the scale is tared,

– the internal signal is stable.

If these conditions are not met, an error message

is displayed.

The weight required for calibration/adjustment is displayed.

Example

Calibrate the scale

Menu code settings: Factory-set codes

Step

Key (or instruction) Display

1. Turn on the scale

[ON/OFF]

2. Tare the scale

t

0.000

ct

3. Begin calibration

[CAL/CF]

+

100.000

Calibration weight

is displayed without

weight unit (here: 100 g)

4. Place the indicated

100.000

calibration weight

on the scale

After calibration, the

+

500.000

ct

calibration weight

is displayed with wt. unit

5. Remove the

0.000

ct

calibration weight

* “Calibration” technically means to determine the difference between

the scale readout and the actual weight on the pan to determine

the accuracy. Adjustment means to bring a scale into the state of accuracy

required for its use. Therefore, “calibration,” as used in this manual,

actually means “adjustment.”

Operating the Scale

Basic Weighing Function

Available Features

– Taring the scale

You can tare the scale within the entire weighing range.

Preparation

l

Turn on the scale: Press [ON/OFF]

l

To change configurations: See the chapter entitled

“Configuring the Scale”

l

To load factory-set configurations:

See “Configuring the Scale,” parameter 9 -1

$

To tare the scale: Press t

Additional Functions

l

To turn off the scale: Press [ON/OFF]

Warmup Time

To deliver exact results, the scale must warm up for

at least 30 minutes after initial connection to AC power.

Only after this time will the scale have reached the

required operating temperature.

Example

Basic weighing

Menu code settings: Factory-set codes

Step

Key (or instruction) Display

1. Turn on the scale

[ON/OFF]

Self-test is performed

2. Place container on scale

+

52.000

ct

(here: 52 ct)

3. Tare the scale

t

+

0.000

ct

4. Place sample in container

+

150.210

ct

on scale (here: 150.210 ct)