Application programs, Counting, Weighing in percent – CITIZEN CARAT BALANCES CT103 User Manual

Page 4

Advertising
background image

4

Application Programs

Counting

Purpose

With the Counting program you can determine the number

of parts that each have approximately equal weight.

Available Features

– Store the current weight value to have it loaded as the

preset reference sample quantity next time you initialize

the Counting application

– The reference sample quantity can be changed in the

operating menu: See “Configuring the Scale”

– The average piece weight is automatically output via the

optional data interface port after initialization, if the menu

code for “Printout with data ID codes” is set

– Press [FUNCTION] to toggle the display between piece

count and weight

Factory Settings

Reference sample quantity: 10 (menu code 3. 3. 2)

Preparation

l

Configure the Counting application in the

operating menu:

See “Configuring the Scale”

Menu code 2. 1. 4

Counting

l

Reference sample quantity:

Menu code 3. 3. 1

5 pcs

Menu code 3. 3. 2

10 pcs

Menu code 3. 3. 3

20 pcs

Menu code 3. 3. 4

50 pcs

Menu code 3. 3. 5

100 pcs

See also “Configuring the Scale”

Example

Determine an unknown piece count; weigh the preset

reference sample quantity

Settings (changes in the factory settings required for

this example):

Menu: Application program: Counting (2 .1. 4)

Reference sample quantity: 20 pcs (3. 3. 3)

Step

Key (or instruction) Display

1. Turn on the scale

[ON/OFF]

2. Tare the scale

t

0.000

c t

3. Display the reference

[FUNCTION]

ref 20

(briefly)

sample quantity

> 2 sec.

(here: 20 pcs)

4. Place the reference

+

16.000

c t

sample quantity

(20 pcs) on the

scale (here: 16 ct)

5. Start the application;

[FUNCTION]

+

20

pcs

if the print format is set

to include data ID codes,

the following is printed

wRef + 0.8000

ct

6. Weigh uncounted parts

+

174

pcs

(here: 174 pcs)

7. Display weight

[FUNCTION]

+

139.200

c t

8. Display quantity

[FUNCTION]

+

174

pcs

9. Unload the scale

0

pcs

10. Delete the reference

value

[CAL/CF] > 2 sec.

11. Repeat the procedure

starting from Step 6,

if desired.

Weighing in Percent

Purpose

This application program allows you to obtain weight

readouts in percent which are in proportion to a reference

weight.

Available Features

– Store the current weight value to have it loaded as the

preset reference percentage next time you initialize the

Weighing in Percent application

– The reference percentage can be changed

in the operating menu: See “Configuring the Scale”

– The reference percentage is automatically

output via the optional data interface port after

initialization, if the menu code for “Printout

with data ID codes” is set

– Press [FUNCTION] to toggle the display between

percentage and weight

Factory Settings

Reference percentage: 10 (menu code 3. 3. 2)

Advertising
This manual is related to the following products: