Calibration procedure – non- sealed- conitnued, Table 3 - calibration weight table – Salter Brecknell C3255 User Manual

Page 13

Advertising
background image

1000 Armstrong Drive. Fairmont, Minnesota 56031 Tel: 507-238-8702 Fax: 507-238-8271 e-mail:

[email protected]

For technical support contact us at 1-800- 458-7062 or fax us at 1-800-368-2042

Page 13 of 18

CALIBRATION PROCEDURE – NON- SEALED- Conitnued

FUNCTION

SCALE RESPONSE

11. Load the appropriate calibration weight.

“XXXXXX”

See Table 3 for valid calibration

weights used, verify weight placed

on platform is with acceptable span

range.

12. Press the

TARE

key to begin calibration

“

-CAL-

“

13. The scale will exit out of calibration mode

and return to normal weighing mode.

“

Scale will display weight used for

calibration

“

If calibration process does not correct any weighing issues, contact the service department for

technical support at 1-800-458-7062.

Table 3 - Calibration Weight Table

“FULL” – Full capacity

Mode (Capacity

KG)

Capacity LB

Calibration

Weight KG

Calibration

Weight LB

Acceptable

Span Range

CS_ _6 (6KG)

12

N/A

N/A

N/A

CS_15 (15KG)

30

10 KG

20 LB

80000-120000

CS_30 (30KG)

60

20 KG

40 LB

80000-120000

CS_60 (60KG)

150

50 KG

100 LB

100000-150000

CS_150

(150KG)

300

100 KG

200 LB

80000-120000

CS_300

(300KG)

600

N/A

N/A

N/A

* Grey Shaded areas indicate capacities only available for Model 3255.

“1 / 2”

– 1 / 2 Capacity

Mode (Capacity

KG)

Capacity LB

Calibration

Weight KG

Calibration

Weight LB

Acceptable

Span Range

CS_ _6 (6KG)

12

N/A

N/A

N/A

CS_15 (15KG)

30

5 KG

10 LB

40000-60000

CS_30 (30KG)

60

10 KG

20 LB

40000-60000

CS_60 (60KG)

150

20 KG

50 LB

320000-60000

CS_150

(150KG)

300

50 KG

100 LB

40000-60000

CS_300

(300KG)

600

N/A

N/A

N/A

* Grey Shaded areas indicate capacities only available for Model 3255.

Advertising