Calibration procedure –sealed - continued – Salter Brecknell C3255 User Manual
Page 17
Advertising

1000 Armstrong Drive. Fairmont, Minnesota 56031 Tel: 507-238-8702 Fax: 507-238-8271 e-mail:
For technical support contact us at 1-800- 458-7062 or fax us at 1-800-368-2042
Page 17 of 18
CALIBRATION PROCEDURE –SEALED - Continued
The calibration button is located on
the display main PDB, underneath
the battery compartment.
A sealing label covers the access
hole. Remove the label and press the
calibration button.
The scale will now be in the
configuration mode. The display will
prompt
“ModE”.
You are now able to configure and
calibrate the scale. Review
Configuration and calibration sections
to procede.