RS-232 data commands for SBI protocol

The RS232 can be set so a bidirectional connection can be established between the indicator and the host. To
establish this connection set parameter P4 to 7, and configure setting P5 (baud) and P6 (parity) to host device.
Commands can then be sent from the host to the indicator using the following commands (ensure the letters
entered are in CAPS) (<CR> - Enter)

Command Action

Response

W<CR>

Takes a reading

Over capacity -
Under capacity -
Zero point error -
Reading (kg or lb)



<LF>^^^^^^^^^u1u2<CR><LF>H1H2H3<CR><ETX>
<LF>_______u1u2<CR><LF>H1H2H3<CR><ETX>
<LF>----------u1u2<CR><LF>H1H2H3<CR><ETX>
<LF><p>w1w2w3w4w5w6<dp>w7u1u2<CR><LF>H1H2H3<CR><ETX>

S<CR>

Prints Status Bytes <LF>H1H2H3<CR><ETX>

Z<CR>

Zeros the scale

<LF>H1H2H3<CR><ETX>

T<CR>

Sets up a tare

<LF>H1H2H3<CR><ETX>

U<CR>

Changes the units

<LF>u1u2<CR><LF>H1H2H3<CR><ETX>

L<CR> Activates

the

hold

function

<LF>H1H2H3<CR><ETX>

X<CR> Switches

off

the

scale

Indicator switches off.

? Unrecognised

command

<LF>?<CR><ETX>


Key Symbols

<LF> Line

feed

<CR> Carriage Return
<ETX> End of text character
<SP> Space
H1H2H3 3 status bytes
<p> Polarity character including minus sign for negative weigh and a space character for positive.
W1-W7 Weight data
<dp> Decimal

point

U1U2: Unit measure, kg, lb or oz

Output Status Bit Meaning:

Bit Byte 1

Byte 2

Byte 3

0

0=Stable 0=Not

Under

Capacity

00=Not defined

1=Unstable

1=Under Capacity

01=Normal working mode

1

0=Not at zero
point

0=Not over
capacity

10=Hold working mode

1=At zero point

1=Over capacity

11=Not defined

2

Always 0

Always 0

0=Gross Weight
1=Net Weight

3

0=eprom OK
1=eprom error

Always 0

Always 0

4

Always 1

Always 1

Always 1

5

Always 1

Always 1

Always 1

6

Always 0

Always 1

Always 0

7

Parity Parity Parity