Spectrum Analyzer / Receiver

•

Use the BNC connectors on the front panel to connect the spectrum

analyzer or EMI receiver.

•

The monitor port which is not being monitored should be terminated

into a coaxial 50-ohm terminator.

Connect the input and output terminals to their proper mains

connections and EUT connections before connecting the monitor port

to the measurement instrumentation; otherwise, power surges may

damage the test instrumentation mixers or attenuators.

•

Removal of a terminator and connection to the BNC receptacle will not

generate power surges.

•

When the power is to be disconnected, remove the coaxial connection

to the measurement instrumentation first to avoid possible damage.

18 |

Installation