Conducted susceptibility, Typical test setup, 0 application – ETS-Lindgren 95242-1 Injection Probes User Manual

Page 17

Advertising
background image

4.0 Application

The principal use of the Bulk Current Injection Probe (BCIP) is for inducing

relatively large RF currents into the signal and power circuits of equipment under

test for conducted susceptibility. A secondary application would be to use the

same probe in a more familiar role as a sensor for measuring weak conducted

RF currents.

Conducted Susceptibility

Conducted susceptibility testing is intended to ensure that RF signals, when

coupled on to interconnecting cables and power supply lines of a device under

test (DUT), will not cause malfunction or degradation of performance. In addition,

this testing can provide an amplitude vs. frequency malfunction signature for the

system which, when compared with the levels of current on the cables in a typical

operating environment, can assist in the determination of adequate safety

margins.

T

YPICAL

T

EST

S

ETUP

Typical conducted susceptibility tests require that all power and interconnecting

cables be tested by subjecting them to the required current or voltage levels,

while monitoring the applied current using a current probe. Usually, a reference

level calibration is performed using a calibration jig with a specified impedance.

This reference curve is then replayed to expose the DUT to a controlled stress

level, while a current probe is used to ensure that a low impedance DUT is not

overstressed.

Some tests may allow the reference calibration to be performed at a

lower level and then scaled up to the required power level when

applied to the DUT.

Entire cables or cable bundles may be tested, or each line may be broken out

and tested individually. Some standards may also require simultaneous injection

onto multiple cable bundles using several injection probes. Absorbing clamps

may be required to isolate peripheral equipment from the DUT, and ensure that

only the DUT is exposed to the required stress level. See the pertinent test

standard for more details.

Application

|

17

Advertising
This manual is related to the following products: